4.15 CPR Handle Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Basic Care™ Bed Service Manual (man336)
Page 4 - 35
4
4.15 CPR Handle Assembly
Tools required:
Phillips head screwdriver
8 mm Allen™
1
wrench
Needle nose pliers
Removal
1. Raise the head section to the highest position.
2. Adjust the bed height to a comfortable working position.
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could result
in personal injury or equipment damage.
3. Disconnect the bed from its power source.
4. Disconnect the battery cable from the control board (refer to procedure 4.7
5. Remove the two screws (A) and two bushings (B) attaching the CPR
handle assembly (C) to the bed frame (D) (see figure 4-15 on page 4-36).
6. Remove the spring (E) from the CPR handle assembly (C).
7. Loosen the screw (F) that is attaching the CPR wire (G) to the bracket (H).
8. Remove the CPR wire (G) from the bracket (H).
9. Remove the CPR handle assembly (C) from the bed frame (D).
1. Allen™ is a trademark of Industrial Fasteners, Inc.