4.1 Control Box Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Basic Care™ Bed Service Manual (man336)
Page 4 - 3
4
4.1
Control Box Assembly
Tools required:
5 mm Allen™
1
wrench
Screwdriver
Removal
1. Raise the head section to its highest position.
2. Adjust the bed height to have access to the cables (A) (see figure 4-1 on
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could result
in personal injury or equipment damage.
3. Disconnect the bed from its power source.
4. Disconnect the cables (A) from the control box assembly (B) (see figure 4-
NOTE:
It may be necessary to use a screwdriver as leverage to disconnect the cables
from their sockets.
5. While supporting the control box assembly (B), remove the three screws
(C) attaching the mounting bracket (D) to the bed frame (E).
6. Remove the control box assembly (B) and the mounting bracket (D) from
the bed frame (E).
1. Allen™ is a trademark of Industrial Fasteners, Inc.