2.10 The Hilow Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Page 2 - 20
Basic Care™ Bed Service Manual (man336)
2.10 The Hilow Function Does Not Operate
1. The hilow function lockout is off.
NOTE:
If the LED for a bed function lockout is on, the function is locked out.
Yes
No
↓
→
Turn the hilow function lockout off. If this corrects the problem,
go to “Final Actions” on page 2-4.
2. Disconnect the bed from its power source and allow the bed to cool for 20
minutes. Connect the bed to its power source. The hilow function does not
operate.
Yes
No
↓
→
Go to “Final Actions” on page 2-4.
3. Call Hill-Rom Technical Support at 1-812-934-1796.