4.5 Trendelenburg and Reverse Trendelenburg Indicators
Chapter 4: Removal, Replacement, and Adjustment Procedures
Basic Care™ Bed Service Manual (man336)
Page 4 - 13
4
3. Lower the bed until it stops moving. The bed is now at 0° Trendelenburg
and Reverse Trendelenburg.
CAUTION:
Use care when removing indicator caps. Failure to do so could damage
the indicator cap.
4. If the indicator cap (A) is installed, carefully remove it.
5. Align the mark on the indicator cap (A) with the 0° marks on the angle
indicator label (B).
6. Put the indicator cap (A) in the lift arms (B).
7. To disengage the Trendelenburg, raise the bed to the highest position.