![Hill-Rom Basic Care P1440 Service Manual Download Page 141](http://html1.mh-extra.com/html/hill-rom/basic-care-p1440/basic-care-p1440_service-manual_2134161141.webp)
Lubrication Requirements
Chapter 6: General Procedures
Basic Care™ Bed Service Manual (man336)
Page 6 - 3
6
When servicing the P.C. board do the following:
• Make sure hands are clean and free of moisture, oily liquids, etc.
• Only handle the P.C. board by its outer edges.
• Do not touch the P.C. board components. Finger contact with the board
surface and/or with its components can leave a deposit that will result in
board (and component) deterioration.
• When working with electronics, wear an appropriate antistatic strap, and
make sure it is correctly grounded.
• Service the removed P.C. board at a static-free workstation that is correctly
grounded.
• For shipping and storage, place the removed P.C. board in an antistatic
protective bag.
Lubrication Requirements
WARNING:
Obey the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.
CAUTION:
Do not use silicone-based lubricants. Equipment damage could occur.
Oilite®
1
bearings and bushings are utilized in several places on the system. By
retaining oil, the pores give a self-lubricating quality to the bearings and
bushings. If any silicone-based lubricant is applied to the bearings and
bushings or anywhere else on the system, this self-lubricating quality is
neutralized.
It is safe to apply the following lubricants to the system (see table 6-1 on page
2-3):
Table 6-1. Lubricants
1. Oilite® is a registered trademark of Beemer Precision, Incorporated.
Part Number
Description
8252 (1400)
Non-detergent SAE30 weight oil (apply to Oilite®
bearings and bushings)
SA3351 (1400)
4 oz lithium grease