4.19 Siderail Control Assembly
Chapter 4: Removal, Replacement, and Adjustment Procedures
Basic Care™ Bed Service Manual (man336)
Page 4 - 45
4
4. Remove the five screws (C) attaching the two halves of the siderail control
assembly (D) and (B).
5. Disconnect the cable (E) from the top socket (F) on the outer half of the
siderail control assembly (D).
6. Disconnect the siderail cable (G) from the bottom socket (H) on the outer
half of the siderail control assembly (D).
7. Remove the two screws (I) attaching the strain relief (J) to the outer half of
the siderail control assembly (F).
8. Gently bend the connector (K) down against the siderail cable (G).
9. Push the connector (K) and the siderail cable (G) out of the outer half of the
siderail control assembly (D).
10. Remove and discard the two halves of the siderail control assembly (D)
and (B).
Replacement
1. Remove and save the center screw (C) from the new outer half of the
siderail control assembly (D).
2. Insert the connector (K) and the siderail cable (G) into the outer half of the
siderail control assembly (D).
3. Connect the connector (K) to the bottom socket (H) on the outer half of the
siderail control assembly (D).
4. Connect the cable (E) into the top socket (F) of the outer half of the siderail
control assembly (D).
5. Install the two halves of the siderail control assembly (D) and (B) around
the post (L) on the siderail (M).
6. Install the five screws (C) to attach the two halves of the siderail control
assembly (D) and (B) to the post (L).
7. Install the inner label (A) to the inner half of the siderail control assembly
(B).
8. Connect the battery cable to the control board (refer to procedure 4.7 on