Initial Actions
Chapter 2: Troubleshooting Procedures
Page 2 - 2
Basic Care™ Bed Service Manual (man336)
Initial Actions
Use Initial Actions to gather data from operators about problems with the
Basic Care™ Bed. Note symptoms or other data about the problem that the
operator describes. This data helps identify the most possible cause.
1. Someone who can explain the problem is available.
Yes
No
↓
→
Go to “Function Checks” on page 2-2.
2. Ask that person to demonstrate or explain the problem. The problem can be
demonstrated.
Yes
No
↓
→
Go to “Function Checks” on page 2-2.
3. The problem is a result of incorrect operator action.
Yes
No
↓
→
Go to “Function Checks” on page 2-2.
4. Instruct the operator to refer to the Basic Care™ Bed User Manual
(usr124). To make sure the Basic Care™ Bed is operating correctly, do the
“Function Checks” on page 2-2.
Function Checks
1. Initial Actions have been done.
Yes
No
↓
→
Go to “Initial Actions” on page 2-2.
2. The bed is connected to the correct power source.
Yes
No
↓
→
Connect the bed to the correct power source. If this corrects the
problem, go to “Final Actions” on page 2-4. Otherwise, go to
step 3.
3. Test all of the bed functions for up and down operation. All of the up and
down bed functions operate correctly.
Yes
No
↓
→
Make sure the lockouts for all of the bed functions are not locked
out. If this corrects the problem, go to “Final Actions” on page
2-4. Otherwise, go to step 4.