Specifications
Chapter 1: Introduction
Page 1 - 6
Basic Care™ Bed Service Manual (man336)
Table 1-8. Guidance and Manufacturer’s Declaration—Electromagnetic
Emissions
Table 1-9. Guidance and Manufacturer’s Declaration—Electromagnetic
Immunity
The Basic Care™ Bed is intended for use in the electromagnetic environment specified
below. The customer or the user of the bed should make sure that it is used in such an envi-
ronment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
RF Emissions
CISPR 11
Group 1
The Basic Care™ Bed uses RF energy only for its
internal functions. Therefore, its RF emissions are
low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The Basic Care™ Bed is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Not
applicable
Voltage
Fluctuations/
Flicker Emissions
IEC 61000-3-3
Not
applicable
The Basic Care™ Bed is intended for use in the electromagnetic environment specified
below. The customer or the user of the bed should make sure that it is used in such an envi-
ronment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.