![Hill-Rom Basic Care P1440 Service Manual Download Page 73](http://html1.mh-extra.com/html/hill-rom/basic-care-p1440/basic-care-p1440_service-manual_2134161073.webp)
4.7 Battery Backup
Chapter 4: Removal, Replacement, and Adjustment Procedures
Basic Care™ Bed Service Manual (man336)
Page 4 - 17
4
5. Make a note of the position of the two batteries (D).
6. Disconnect the battery cable (E) from the control board (F).
7. Disconnect the battery cable (E) and the fuse assembly (G) from the two
batteries (D).
CAUTION:
Dispose of batteries according to your local regulations. Failure to
dispose of batteries correctly may damage the environment.
8. Dispose of the two batteries (D) according to your local regulations.
Replacement
1. Do the removal procedure in the reverse order.