4.11 Hilow Motor
Chapter 4: Removal, Replacement, and Adjustment Procedures
Basic Care™ Bed Service Manual (man336)
Page 4 - 25
4
4.11 Hilow Motor
Tools required:
10 mm Allen™
1
wrench
13 mm socket
Needle nose pliers
Wire cutters
Bed supports
Removal
WARNING:
Do not work under an unsupported load. Install appropriate temporary
supports. Failure to do so could result in personal injury or equipment
damage.
1. Put bed supports under the upper bed frame.
2. Disconnect the battery cable from the control board (refer to procedure 4.7
3. Raise the head section (A) to its highest position (see figure 4-11 on page
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could result
in personal injury or equipment damage.
4. Disconnect the bed from its power source.
5. Make a note showing the cable routing from the hilow motor (B) to the
control box assembly (C).
CAUTION:
Use caution when cutting wire ties. Failure to do so could result in cable
damage.
6. Carefully cut all of the wires ties attaching the hilow motor cable (D) to the
bed frame (E).
7. Disconnect the hilow motor cable (D) from the control box assembly (C).
1. Allen™ is a trademark of Industrial Fasteners, Inc.