6.1 Basic Care™ Bed—Sleep Mode
Chapter 6: General Procedures
Page 6 - 8
Basic Care™ Bed Service Manual (man336)
6.1
Basic Care™ Bed—Sleep Mode
If the bed will be disconnected from its AC power source for an extended
period of time, do this procedure to conserve battery power.
Tools required:
Phillips head screwdriver
1. Raise the head section (A) to its full up position.
2. Lower the bed to have access to the cover (B) on the control box assembly
(C) (see figure 6-1 on page 6-8).
Figure 6-1. Control Box
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could result
in personal injury or equipment damage.
3. Disconnect the bed from its power source.
4. Remove the six screws (D) attaching the cover (B) to the control box (C).
5. Disconnect the battery cable (E) from the control board (F).