2.5 The Knee Up Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Basic Care™ Bed Service Manual (man336)
Page 2 - 11
2
2.5
The Knee Up Function Does Not Operate
1. The knee up function does not operate when the knee up switch is pressed
from either the left siderail or the right siderail or from the foot end control
panel.
Yes
No
↓
→
2. The knee function lockout is off.
NOTE:
If the LED for a bed function lockout is on, the function is locked out.
Yes
No
↓
→
Turn the knee function lockout off. If this corrects the problem,
go to “Final Actions” on page 2-4.
3. Tightly seat the connector for the knee motor into its socket on the control
box. The knee up function does not operate.
Yes
No
↓
→
Go to “Final Actions” on page 2-4.
4. Replace the knee motor. The knee up function does not operate.
Yes
No
↓
→
Go to “Final Actions” on page 2-4.
5. Call Hill-Rom Technical Support at 1-812-934-1796.
6. The knee up function operates from the foot end control panel.
Yes
No
↓
→
Replace the foot end control panel. If this corrects the problem,
go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 1-812-934-1796.
7. The knee up function only operates from one of the siderails.
Yes
No
↓
→
Tightly seat both of the siderail connectors into their sockets on
the control box. If this corrects the problem, go to “Final
Actions” on page 2-4. Otherwise, call Hill-Rom Technical
Support at 1-812-934-1796.