![Hill-Rom Basic Care P1440 Service Manual Download Page 47](http://html1.mh-extra.com/html/hill-rom/basic-care-p1440/basic-care-p1440_service-manual_2134161047.webp)
2.11 The Automatic Contour Up Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Basic Care™ Bed Service Manual (man336)
Page 2 - 21
2
2.11 The Automatic Contour Up Function Does Not Operate
1. The head section does not rise when the control for the auto contour up is
pressed.
Yes
No
↓
→
2. The head function lockout is off.
NOTE:
If the LED for a bed function lockout is on, the function is locked out.
Yes
No
↓
→
Turn the head function lockout off. If this corrects the problem,
go to “Final Actions” on page 2-4.
3. Do the procedure on page 2-6. The head section rises when the control for
the auto contour up is pressed.
Yes
No
↓
→
Call Hill-Rom Technical Support at 1-812-934-1796.
4. The knee section does not rise when the automatic contour up switch is
pressed.
Yes
No
↓
→
Turn the knee function lockout off. If this corrects the problem,
go to “Final Actions” on page 2-4. Otherwise go to step 5.
NOTE:
If the LED for a bed function lockout is on, the function is locked out.
5. Do the procedure on page 2-11. If this corrects the problem, go to “Final
Actions” on page 2-4. Otherwise, call Hill-Rom Technical Support at
1-812-934-1796.