Component Handling
Chapter 6: General Procedures
Page 6 - 2
Basic Care™ Bed Service Manual (man336)
Steam Cleaning
Do not use any steam cleaning device on the Basic Care™ Bed. Excessive
moisture can damage mechanisms and components in this unit.
Hard to Clean Spots
To remove difficult spots or stains, use standard household cleaners and a soft
bristle brush. To loosen heavy, dried-on soil or excreta, it may be necessary to
first saturate the spot.
Disinfecting
Dilute disinfectants and germicides as specified on the manufacturer’s label.
Component Handling
CAUTION:
Make sure that your hands are clean, and only handle the P.C. boards
by their edges to prevent component damage.
CAUTION:
Wear an antistatic strap when handling electronic components. Failure
to do so could result in component damage.
CAUTION:
Put the removed P.C. boards in antistatic protective bags for shipping
and storage. Equipment damage can occur.
P.C. Board
Be careful with the P.C. board when servicing it, or the following may occur:
• P.C. board damage
• Shortened P.C. board life
• Unit malfunctions