6.2 CPR Cable Adjustment
Chapter 6: General Procedures
Basic Care™ Bed Service Manual (man336)
Page 6 - 11
6
Figure 6-2. CPR Adjustment
2. If the head section (A) is lowering too quickly when the CPR handle (B) is
operated, do the following:
a. Loosen the screw (C) that is attaching the CPR wire (D) to the bracket
(E).
b. Allow a small part of the CPR wire (D) to go into the cable (F).
c. While holding the CPR wire (D) tightly, tighten the screw (C) to attach
the wire (D) to the bracket (E).
NOTE:
While the CPR wire is being held tightly, the head section will be lowering
slowly.
d. Operate the CPR handle (B) to make sure the head section (A) lowers
at the correct rate.
e. If necessary repeat step 1 or step 2 until the head section (A) lowers at
the correct speed.