4.7 Battery Backup
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 16
Basic Care™ Bed Service Manual (man336)
4.7
Battery Backup
Tools required:
Phillips head screwdriver
Needle nose pliers
Removal
1. Raise the head section to the highest position.
2. Adjust the bed height to have access to the cover (A) (see figure 4-7 on
page 4-16).
Figure 4-7. Control Box Assembly
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could result
in personal injury or equipment damage.
3. Disconnect the bed from its power source.
4. Remove the six screws (C) attaching the cover (A) to the control box
assembly (B).