6.2 CPR Cable Adjustment
Chapter 6: General Procedures
Page 6 - 10
Basic Care™ Bed Service Manual (man336)
6.2
CPR Cable Adjustment
Tools required:
Needle nose pliers
Small adjustable wrench
1. If the head section (A) is not lowering quickly enough when operating the
CPR handle (B), do the following (see figure 6-2 on page 6-11):
a. Loosen the screw (C) that is attaching the CPR wire (D) to the bracket
(E).
b. Pull a small part of the CPR wire (D) out of the cable (F).
c. While holding the CPR wire (D) tightly, tighten the screw (C) to attach
the wire (D) to the bracket (E).
NOTE:
While the CPR wire is being held tightly, the head section will be lowering
slowly.
d. Operate the CPR handle (B) to make sure the head section (A) lowers
at the correct rate.
e. If the head section (A) is not lowering quickly enough when the CPR
handle (B) is operated, repeat step 1.
f.
If the head section (A) is lowering too quickly when the CPR handle
(B) is operated, go to step 2.