4.9 CPR Cable
Chapter 4: Removal, Replacement, and Adjustment Procedures
Page 4 - 20
Basic Care™ Bed Service Manual (man336)
4.9
CPR Cable
Tools required:
Phillips head screwdriver
4 mm Allen™ wrench
6 mm wrench
8 mm wrench
Needle nose pliers
Wire cutters
Removal
1. Disconnect the battery cable from the control board (refer to procedure 4.7
2. Raise the bed to the highest position.
3. Raise the head section to the highest position.
SHOCK HAZARD:
Disconnect the bed from its power source. Failure to do so could result
in personal injury or equipment damage.
4. Disconnect the bed from its power source.
5. Make a note showing the cable routing from the CPR handle assembly (A)
to the head motor (B).
CAUTION:
Use caution when cutting wire ties. Failure to do so could result in cable
damage.
6. Carefully cut all of the wires ties attaching the CPR cable (C) to the bed
frame (D).
7. Loosen the screw (E) that is attaching the CPR wire (F) to the bracket (G).
8. Remove the nut (H) attaching the CPR cable (C) to the CPR handle
assembly (A).
9. Remove the CPR cable (C) from the CPR handle assembly (A).
10. Remove the screw (I), nut (J), and two washers (K) attaching the CPR wire
(F) to the lever (L).