Section 8-2 - External Component Replacement Procedures
6) Return the HVPS as described in the
HVPS Installation Procedure on page 8 - 48
7) Return the BEP as described in the
BEP Installation Procedure on page 8 - 42
.
8) Return the TR32 and RFI Boards as described in the
TR32 and RFI Boards Installation Procedure
9) Return the P&M Board as described in the
P&M (Probe and MUX) Board Installation Procedure on
10) Return the Control Panel and Keyboard as described in the
Control Panel and Keyboard Installation
.
11) Return the hard disk as described in the
Hard Disk Installation Procedure on page 8 - 8
.
12) Return the bearing handle as described in the
Bearing Handle Installation Procedure on page 8 - 4
.
13.)Re-install the battery as described in the
Battery Installation Procedure on page 8 - 6
14.)Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
- only if the Vivid
i
n/ Vivid q N system
is currently mounted on the SafeLock cart
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
-
Calibration Tests on page 7 - 25
-