GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
Chapter 7 - Diagnostics/Troubleshooting
7-77
7) When finished reviewing the first report, trackball to the
Exit
button and press
Set
to close the View
Report window.
8) Select the second report (
BPF1
); the report is highlighted in the
Reports
list, and the corresponding
results are shown in the Report window below, as shown in
9.) Use the tools provided to scroll through the channels and view the results, as previously described.
10) When done, trackball to the
Close
button and press
Set
to close the View Report window.
11) Select the third report (
BPF2
); the report is highlighted in the
Reports
list, and the corresponding
results are shown in the Report window below, as shown in
Figure 7-49 View Test Log - ACW BPF Test - BPF1