Chapter 4 - General Procedures and Functional Checks
4-13
4-2-5-2
Install MO Disk in MO Drive
1.) Preparations before installing an MO disk in the MOD Drive:
a.) The unit must be turned ON and "up and running" before using the MOD Drive
b.) Verify the MO disk for loose hardware or damaged labels which could jam inside the MOD
Drive.
c.) Ensure that the slide switch in one corner of the disk is set so that the disk is write enabled
(disk hole closed).
2.) Insert the disk into the MO Disk Drive with the label for the side you intend to use, facing up.
4-2-5-3
Eject and Remove MO Disk from MOD Drive
In some cases, the MO Disk is ejected automatically. If not automatically ejected, use the manual
ejection method as described in the following steps:
1.) Select the
Eject
button on the screen
OR
Press <
Alt> + E
on the alphanumeric keyboard to display the Eject Device menu.
2.) Select
5 1/4 MO Disk
in the Eject Device menu to eject the MO Disk.
3.) Remove the ejected MO Disk from the MO Drive and place it in its cover box for storage in a safe
place.
4-2-5-4
Connecting External USB Hard Disk to USB Port
The External USB Hard Disk can be connected to either of the USB ports (
upper
or
lower
) on the rear
connectors panel.
One must use the additional power source (supplied with the External USB Hard Disk) to activate the
External USB Hard Disk.
WARNING
Do not attempt to use a different type of mass storage device other than the
External USB Hard Disk supplied with the system. The ultrasound system is an extremely
sensitive and complex medical system. Any unauthorized peripherals may cause system failure
or damage!
Refer to the
External USB Hard Disk Installation Procedure
section, page 8-151 for instructions on initial
configuration and connecting the External USB Hard Disk to the Vivid
i
n/Vivid q N portable ultrasound
scanner scanner.
After connecting the External USB Hard Disk to the system, perform the functional checks described in
the
External USB Hard Disk Test on page 4-32
.
NOTICE
Never move the unit with a MO disk in the Magneto Optical Disk Drive because the drive actuator
will not be locked and the Magneto Optical Disk Drive could break.
NOTICE
Do not eject the MO Disk using the button on front of the MO drive.
Figure 4-6 Eject Device Menu