Chapter 8 - Replacement Procedures
8-19
8-2-5
Rear Cover & Latch Assembly Replacement Procedure
NOTE:
This section describes the replacement procedure for Rear Cover & Latch Assembly - supplied as part
of the Cover Kit (Vivid i n systems Part No. S2423361; Vivid q N systems Part No. S2423362).
8-2-5-1
Tools
None required.
8-2-5-2
Time Required
30 minutes
8-2-5-3
Preparations
Shut down the Vivid
i
n/ Vivid q N ultrasound unit, as described in
8-2-5-4
Rear Cover & Latch Assembly Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid
i
n/Vivid q N portable ultrasound scanner.
2) Remove the Bearing Handle as described in the
Bearing Handle Removal Procedure on page 8 - 3
.
3) Remove the Control Panel and Keyboard as described in the
Control Panel and Keyboard Removal
4) Remove the LCD Display Frame as described in the
LCD Display Frame Removal Procedure on
5)
Remove the LCD Display; see the
LCD Display Removal Procedure on page 8 - 53
.
6.) Loosen and remove the two screws on each of the
left
and
right
hinges (attached to the Rear Cover
and Latch Assembly), as shown in
7.) Lift the Rear Cover and Latch Assembly away from the Bottom Assembly and remove completely
from the system.
8-2-5-5
Rear Cover & Latch Assembly Installation Procedure
1.) Return the Rear Cover and Latch Assembly to the system and place it in position on the Bottom
Assembly, carefully checking that the
left
and
right
hinges are properly aligned.
2.) Return and fasten the two screws on each of the
left
and
right
hinges (previously removed - refer
3)
Return the LCD Display; see the
LCD Display Installation Procedure on page 8 - 55
4)
Return the LCD Display Frame as described in the
LCD Display Frame Installation Procedure on
Figure 8-23 Removing the Hinge Screws