Chapter 3 - System Setup
3-19
3-5-4
Confirming SafeLock Cart Voltage Configuration
1.) Check the safety rating plate located at the rear of the SafeLock Cart, above AC input connector.
2.) Remove the bottom cover (rear) from the SafeLock Cart to provide access to the jumper on the AC
unit. See
3.) Make sure that the jumper is set to either 100 to 120 V AC or to 220 - 240 V AC, in accordance with
the voltage rating shown on the safety rating plate (step 1).
3-5-4-1
Connection and Usage of the optional Modo Cart
Please refer to the “Vivid i n Cart, User Manual – Supplement to Vivid i n User manual”,
Direction Number FL092096.
3-5-5
System Voltage Confirmation
3-5-5-1
System Voltage Settings
Verify that the
Vivid
i
n and Vivid q N is set to the correct voltage.
The
Vivid
i
n and Vivid q N voltage settings are found on the base (underside) of the system.
For illustrations of the label with the voltage setting, see:
•
Table 1-6 "Main Label for Vivid q N" on page 1-12
•
Table 1-7 "Main Label for Vivid i n" on page 1-12
Figure 3-12 Jumper on AC Unit
WARNING
WARNING
CONNECTING THE VIVID
I n/ Vivid q N
TO THE WRONG VOLTAGE LEVEL WILL MOST
LIKELY DESTROY THE SCANNER.