Section 8-2 - External Component Replacement Procedures
8-2-3-5
Hard Disk Installation Procedure
NOTE:
When replacing the BEP or Hard Disk, one must either obtain new access codes that will fit the
system ID which is programmed on the BEP, or program the BEP system ID to fit the access codes
already registered in the Hard Disk. Only qualified personnel should perform these tasks.
NOTE:
When performing the following installation procedure steps, it may be helpful to review the appropriate
movie clips previously referenced in the
.
1) Place the new hard disk in position in the hard disk module holder - see
.
2) Return and fasten the two screws (previously removed) to secure the Hard Disk in position - refer
.
Note:
When a new hard disk arrives from the manufacturer, usually it is partitioned and ready for
software installation.
Refer to
Software Loading on page 8 - 97
for software installation instructions .
Note:
IMPORTANT - if
both
the Hard Disk and Back End are being replaced simultaneously, it is
necessary to re-configure the LCD type. Refer to the instructions provided in
page 8-56 through
3.) Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
- only if the Vivid
i
n/ Vivid q N system
is currently mounted on the SafeLock cart
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
-
Calibration Tests on page 7 - 25
-
Full System Test on page 7 - 10