GE H
EALTHCARE
D
IRECTION
FQ091013, R
EVISION
1
V
IVID
I
N
AND
V
IVID
Q
N S
ERVICE
M
ANUAL
Chapter 8 - Replacement Procedures
8-57
11) Select the LCD screen type that is installed on the system, as shown in the example below.
12) Click the
Burn
button.
The system is now updated with the new LCD configuration.
13) For the final stage of the LCD Display installation procedure, proceed to calibrate the
newly-installed LCD Display by making any adjustments required to the
Brightness
,
Contrast
and/or
Blue Tint,
as described in the
LCD Display Adjustments and Calibration on page 6 - 2
14.)Proceed to perform the following functionality tests:
-
SafeLock Cart - Grounding Continuity on page 10 - 23
- only if the Vivid
i
n/ Vivid q N system
is currently mounted on the SafeLock cart
-
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
-
Full System Test on page 7 - 10
Figure 8-67 Selecting LCD Type Installed on the System
Summary of Contents for Vivid in
Page 2: ......
Page 30: ...GE HEALTHCARE DIRECTION FQ091013 REVISION 1 VIVID I N AND VIVID Q N SERVICE MANUAL xxviii...
Page 739: ......