Section 8-2 - External Component Replacement Procedures
the handle in position), by turning 1/4 turn in a
counterclockwise
direction (
).
3) Using both hands, pull the Bearing Handle slightly apart to release it from the system.
4.) Remove the Bearing Handle.
8-2-1-5
Bearing Handle Installation Procedure
1) Place a new Bearing Handle in position, using both hands to pull the Bearing Handle slightly apart
in order to slot the ends into the handle holes on the Vivid
i
n casing.
2) Carefully fasten each captive screw on either side of the Bearing Handle to secure the handle in
position, by turning 1/4 turn in a
clockwise
direction.
Note:
When tightening the captive screws, do not use excessive force
do not apply more than
1/4 turn
; overtightening will damage the screw thread.
3) Proceed to perform the
Bearing Handle Test on page 4 - 19
.
Figure 8-2 Releasing the Bearing Handle Captive Screw
Figure 8-3 Bearing Handle Removed