Chapter 6 - Service adjustments
6 - 13
6-7-2
Adjusting the XY Lock
Use a 3 mm hex tool to adjust the XY Lock release for optimal lock and minimal slack in transport mode.
•
Rotating the screw clockwise will tighten the Lock and will reduce slack, but will also require more
force for engaging the Lock.
•
Rotating counterclockwise will open the lock (more slack).
Follow the steps below to adjust the XY Lock:
1.) Locate the hex screw behind the XY Assembly. This is the adjustment screw.
NOTE:
When turning the hex screw, do not exceed half turn increments. Overtightening will prevent the
console from locking into place and too loose will make the console loose.
2.) Use a 3 mm hex tool to rotate the adjustment screw clockwise to tighten the XY Lock.
For optimal adjustment, rotate the adjustment screw 1/4 turn and test the Lock function. Repeat
procedure if needed.
Rotating the screw counterclockwise will loosen the lock.
Too tight will reduce the slack and make it harder to engage the Lock. Make the adjustments in 1/4 turn
increments and test the lock until the optimum adjustment is achieved.
Figure 6-12 Adjusting the XY Locking mechanism
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......