Chapter 8 - Replacement procedures
8 - 293
8-12-1-1
Main Power Supply removal procedure
Follow these steps to remove the Main Power Supply:
1.) Disconnect the cables on top of the power supply.
2.) Remove the three screws that secure the Main Power Supply.
3.) Ease the Main Power Supply away from the fang bracket.
4.) Pull/slide the entire Main Power Supply backwards until the connectors to the Front End Processor
are unseated.
5.) Lift the Main Power Supply away.
Figure 8-245 Main Power Supply
Figure 8-246 Main Power Supply cables
Main Power Supply
SCREW PLACEMENT
AC OUTPUT TO
PERIPHERALS,
PRINTER
FANG
BRACKET
POWER TO BEP
USB
SCREW