Chapter 8 - Replacement procedures
8 - 117
8-6-11-2
High Voltage Inverter Board with Cable removal procedure
1.) Disconnect the cables to the High Voltage Inverter Board.
2.) Remove the two fixing screws.
3.) Remove the High Voltage Inverter Board with Cable.
8-6-11-3
High Voltage Inverter Board with Cable installation procedure
1.) Align the board’s fixing holes to the respective holes on the Main Controller Board.
2.) Install the two fixing screws. (Size: M 2.5)
3.) Plug in the cable to the connector on the Main Controller Board.
4.) Install the Operator Panel, Upper.
5.) Install the Control Knobs.
Figure 8-131 Remove cables
Figure 8-132 Remove screws