Chapter 8 - Replacement procedures
8 - 15
8-4-9-3 Initiate software reload from the Recovery Console (cont’d)
If, accepted, the screen below is displayed.
6.) Press the
Enter
key on the keyboard. The software reload starts.
NOTE:
Time to complete the software reload is typically 15 to 25 minutes.
-
When the
System Software reload
is done, the VIVID E9 reboots.
-
After the reboot, the VIVID E9 starts to load the Application Software. This is done without any
interactions. The TCP/IP settings, Computer Name etc. are restored.
-
The VIVID E9 shuts down.
Figure 8-4 Reload System Software