GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 4 - General procedures and Functional checks
4 - 13
4-2-3-6
Moving the Top Console up or down when Power is OFF
The Z mechanism can be manually repositioned (moved up or down) in the event the drive gear is
disconnected or has failed, or if the mains power is turned off or disconnected, or if it is a mains power
failure.
•
Move the handle a few millimeters to the right side, to release the Z Mechanism (disengage the
motor drive unit), and at the same time move the Top Console up or down, as required.
Figure 4-12 Up/Down Release Handle
WARNING
WARNING
OPERATOR CONSOLE CAN DROP UNEXPECTEDLY WHEN THE MECHANISM IS
RELEASED.
MOVE HANDLE (ABOVE THE ARROW IN
THE ILLUSTRATION) TO THE RIGHT TO
RELEASE THE Z MECHANISM