Section 8-6 - Top Console Parts Replacement
8-6-14-2
Options Holder removal procedure
Follow these steps to remove the Options Holder:
1.) From underneath the Upper Console, remove the 3 screws securing the Options Holder.
2.) Remove the Options Holder from the Upper Console.
8-6-14-3
Options Holder installation procedure
Follow these steps to install the Options Holder:
1.) Position the Options Holder in place.
2.) Install the 3 screws securing the Options Holder.
3.) Install the Storage Tray to the Options Holder.
8-6-15
Knobs for Encoders and Slidepots replacement
8-6-15-1
Tools
No tools are needed for this procedure.
8-6-15-2
Knobs for Encoders and Slidepots removal procedure
NOTE:
If you are going to remove the Upper Operator Panel, you only need to remove the five knobs
below the Touch Screen.
1.) Pull the knobs one by one until all knobs have been removed.
2.) Store the knobs in a clean place.
Figure 8-135 Screws, beneath upper console
WARNING
WARNING
DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS.
1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED
ESD CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (NEAR THE
POWER CONNECTOR).
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......