5-11-3-2
Location in the Unit
The I/O board is located inside the BEP with some connectors available on the rear of the system.
5-11-3-3
Input signals
•
Supply voltage of 5V and 12V from BEP (and BEP Power Card).
•
5V STDBY from the BEP (and BEP Power Card) for OPIO (pass through).
•
PWR_ON signal from the Operator Panel to the BEP’s Motherboard (pass through).
•
Two USB ports for OP Panel (pass through)
•
Two USB ports for the Rear IO
•
One USB port for internal B/W printer
•
One USB port for XYZ Controller
•
Some unused USB ports
•
Doppler audio (orginating on the GFI board)
•
Doppler audio mixed with Windows system sounds from BEP’s motherboard
•
DVI-I input from the BEP (from the Graphics Adapter or onboard video controller, or from the DVR
Board.
5-11-3-4
Output signals
•
12V DC and 5V DC for Patient IO, DVD drive and HDD.
•
5V STDBY from the BEP power supply card for OPIO (pass through)
•
USB ports
•
Audio to the OP Panel for the speakers.
•
External Audio outputs
•
Audio to the woofer (center speaker)
•
DVI-D output for Main Monitor (Digital Video)
•
DVI-I output for External Monitor (only the Digital Video Signal is included)
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......