Section 10-5 - System maintenance
10-5-2-2
Peripheral/option checks
If any peripherals or options are not part of the VIVID E9 configuration, the check can be omitted.
Refer to the User Manual for a list of approved peripherals/options.
10-5-3
Input Power
10-5-3-1
Mains cable inspection
Table 10-4 GE Approved Peripheral/Hardware Option Functional Checks
Step
Item
Description
1.
Media
Verify media drive(s) read/write properlty. Clean if necessary.
2.
B/W Printer
Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
3.
Color Printer
Verify hardcopy output of the Color video page printer. Clean heads and covers if necessary.
4.
DICOM
Verify that DICOM is functioning properly. Send an image to a DICOM device.
5.
ECG
Verify basic operation with customer
6.
Footswitch
Verify that the footswitch is functioning as programed.
Clean as necessary.
Table 10-5 Mains Cable Inspection
Step
Item
Description
1.
Unplug Cord
Disconnect the mains cable from the wall and the VIVID E9.
2.
Inspect
Inspect it and its connectors for damage of any kinds.
3.
Verify
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that
no strands may cause a short circuit.
4.
Verify
Inlet connector retainer is functional.
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......