GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 301
8-13-2-4
BEP5 I/O Board Assembly removal
Follow these steps to remove the BEP5 I/O Board Assembly:
1.) Disconnect the cables inside the BEP that connect to the I/O.
-
Flat cable from the Video (Graphics) board.
2.) Remove the five screws connecting the I/O to the BEP.
3.) Reach inside the BEP frame to pull the I/O Board Assembly out of the BEP.
4.) Set the I/O Board
Assembly aside.
Figure 8-250 BEP5 I/O Board Assembly’s screw placement
Figure 8-251 Remove the BEP5 I/O Board Assembly
SCREWS