GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
4 - 12
Section 4-2 - General procedures
4-2-3-5
Manually releasing the XY Lock
Even if the power is OFF, it is possible to manually release the console’s XY-mechanism (frog leg).
The release point is located on the rear of the XY Mechanism (Frog Leg).
1.) Insert a small tipped screwdriver or eq. into the release point and push until a “click” is heard. Some
force may be required if the lock adjustment is tight adjusted.
2.) Move the console to the wanted position.
Figure 4-10 XY mechanism release
Figure 4-11 Manually releasing the XY Lock
XY RELEASE
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......