Chapter 8 - Replacement procedures
8 - 197
8-8-4-3
Filter removal procedure
1.) Thread a long screwdriver through the extending cloth loop (tab).
2.) Pull down the screwdriver to release the filter from the unit.
3.) Remove the filter.
8-8-4-4
Filter Cleaning
You may either use a vacuum cleaner to vacuum the filter, or if needed, clean the filter by rinsing with
water.
After cleaning the filter by rinsing with water, allow it to dry completely before re-installing.
When dry, install the filter as described next.
8-8-4-5
Filter installation procedure
•
Install the filter at the bottom of the VIVID E9. Align it in position (see: the tab in
). The
magnets will help to position it correct.
Figure 8-201 Release of the bottom air filter
CAUTION
DO NOT operate the unit without the air filters in place.
CAUTION
Allow the air filter to dry thoroughly before re-installing it in the unit.
Figure 8-202 VIVID E9 seen from below
LINE UP THE FILTER SO THAT THE MAGNETS
ATTACH IN PLACE ON THE BOTTOM OF THE
VIVID E9.
WITHOUT FILTER
WITH FILTER