Chapter 6 - Service adjustments
6 - 3
6-2-5
Advanced LCD adjustments
NOTE:
Do not adjust these settings unless necessary!
To get access to the advanced adjustments, press the
Mode
button for
more than 10 seconds
. This
will display the Advanced Menu on the screen.
The
Advanced Menu
has the following choices:
6-2-5-1
Review Test Patterns
Through the Touch Panel, access Utilities / Test Patterns to review the test patterns.
Figure 6-2 LCD Controls Advanced Menu
Table 6-1 Advanced Menu
MENU
PARAMETER
DEFAULT
SETTING
Screen
Smoothing: 1 / 2 / 3 / 4 / 5
3
Return
-
Color
Brightness: 0-100%
89%
Contrast: 0-100%
100%
Temperature: Custom / 9000K / 11000K / 13000K / 15000K
9000K
Gamma: 2.0/2.1/2.2 /2.3/ 2.4 /2.5/ 2.6
2.4
Gain: R/G/B 0-100%
Adjusted value at the
factory
Reset
-
Return
-
Other
Screen Size: FullScreen / Enlarged / Normal
FullScreen
Menu Position
Center
Reset
-
Return
-
Information
Resolution, fH/fV
-
Model name, GPN No. (PSP No.), Serial No. , Using time
-
Language
Japanese, English, German, French, Spanish, Italian, Swedish
English
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......