Chapter 8 - Replacement procedures
8 - 149
8-6-24-3
Assemble the A/N Keyboard and the A/N Keyboard Enclosure
1.) If not already done, connect the Ground Cable to the A/N Keyboard, using the current screw and
two additional washers (tooth, 3.2mm). Use one washer under and one washer over the cable shoe.
(Torque: 2.3 Nm.)
2.) Install the A/N keyboard into the A/N Keyboard Bottom Enclosure. First position the keyboard under
the front clips (A) then press the rear side of the keyboard under the rear clips (B).
Figure 8-158 Install Ground Cable
Figure 8-159 Install A/N keyboard into the A/N Keyboard Bottom Enclosure
B
A