GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 134
Section 8-6 - Top Console Parts Replacement
8-6-19-3
Lower Switch Board with Elastomer installation
1.) Verify that all switch cups are in their correct positions.
2.) Position the Lower Switch Board with Elastomer in its correct position.
3.) Install the 26 fixing screws.
4.) Connect the cable to the Encoder Board.
5.) Install the Trackball.
6.) Install the Operator Panel, Lower.
7.) Install the Operator Panel, Upper.