Section 8-6 - Top Console Parts Replacement
8-6-8-2 Remove the Frame w/LCD and Touch Screen (cont’d)
4.) On the Bulkhead Bracket, loosen the ESD wire from the Lower Panel.
5.) Pull and lift the Operator Panel assembly up and away. Be careful to not destroy the fingers on each
side of the Operator Panel, Lower.
6.) Store it on an ESD safe place.
Figure 8-139 OP Grounding
Figure 8-140 Remove OP Panel Lower assembly
YELLOW/GREEN GROUND WIRE FROM
LOWER OP HERE
(NOT ILLUSTRATED)
FINGER