4 - 34
Section 4-2 - General procedures
4-2-9-5
Remove the Trackball
Follow these steps to remove the Trackball:
1.) Unplug the cable connectors from the Trackball.
2.) Use the Hex key to remove the two fixing screws with shims.
3.) Remove the Trackball and the Fixing Ring.
Figure 4-31 Trackball with fixing screws
Figure 4-32 Trackball removed