GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 294
Section 8-12 - Main Power Supply replacement
8-12-1-2
Main Power Supply installation procedure
Follow these steps to install the Main Power Supply:
1.) Place the Main Power Supply base on the VIVID E9 frame, and then tilt the top of the Main Power
Supply toward the rear of the Card Cage until the Main Power Supply is vertical.
2.) Slide the Main Power Supply forward to seat the connectors on the Card Cage.
3.) Engage the tab onto the Fang.
4.) Install the three screws.
5.) Connect the cables at the top of the Main Power Supply.
6.) Install the Rear Cover.
7.) Install the Filter Cover.
8.) Install the Top Cover.
9.) Install both Side Covers.
Figure 8-247 Main Power Supply installation
CONNECTORS TO
BACKPLANE IN THE
CARD CAGE.