Chapter 10 - Care & maintenance
10 - 11
3.
Control Panel
Clean the Control Panel on a weekly basis.
ONLY use the following cleaners on the VIVID E9 Operator Panel:
•
Palmolive Dishwashing Liquid (manufactured by Colgate-Palmolive)
•
Sani Wipes Alcohol-free (manufactured by Micorgen Inc.)
•
T-Spray II (manufactured by Pharmaceutical Innovations, Inc.)
To clean the operator control panel:
•
Turn off the power to the VIVID E9.
•
Moisten a soft, non-abrasive folded cloth or sponge with a mild, general purpose, non-abrasive soap and water
solution or general purpose disinfectant.
Do not use any solution containing abrasive powder or strong chemicals such as, acid or alkaline.
•
Squeeze excess liquid from the cloth/sponge.
•
Gently wipe the surface of the Control Panel.
•
Use a cotton swab to clean around keys or controls.
•
Use a toothpick to remove solids from between keys and controls.
•
Rinse the cloth/sponge with clean running water and wipe the Control Panel again.
•
Use a dry, soft, lint-free cloth to dry the Control Panel. Wait for the Control Panel surface to dry completely.
4.
Keyboard
•
Clean the keyboard as described (above) for cleaning the Control Panel.
NOTE: In the event that disinfection is required or any stubborn stains remain, absorb a small quantity of isopropyl
rubbing alcohol on a soft, dust-free cloth.
•
Wipe the surface of the keycaps with the cloth, making sure that no liquid drips on or between the keys.
•
Allow to dry.
5.
Trackball
•
For cleaning instructions, ref chapter 4.
6.
Probe Holders
Clean the probe holders with warm water and a damp cloth to remove all traces of gel.
(Soaking may be required to remove excess gel).
7.
System Cabinet
On a weekly basis, moisten a soft, non-abrasive folded cloth or spronge with a mild, general purpose, non-abrasive soap
and water solution.
Do not use any solution containing abrasive powder or strong chemicals such as acid or alkaline.
Squeeze excess liquid from the cloth/sponge, then wipe down the top, front, back and both sides of the unit.
Do not spray any liquid directly onto the unit!
•
Rinse the cloth/sponge with clean running water and wipe the unit surfaces again.
•
Use a dry, soft, lint-free cloth to dry the unit surfaces.
•
Wait for the unit surfaces to dry completely.
NOTE! In the event that disinfection is required or any stubborn stains remain, remove them with a soft, dust-free cloth
on which a small quantity of isopropyl rubbing alcohol has been absorbed, as described above for cleaning the
Keyboard.
8.
DVD - CDRW
Drive
•
Clean the drive head and media with the vendor-supplied cleaning kit.
•
Advise the user to repeat this often, to prevent future problems.
•
CDs must be stored away from dust and cigarette smoke. Do not use alcohol or benzene to clean the CD drive.
9.
Peripherals
Clean the peripherals in accordance with the respective manufacturer’s directions.
10.
Air Filters
Clean the unit’s air filters to ensure that a clogged filter does not cause the unit to overheat and reduce system
performance and reliability.
A message is displayed on screen on a regular basis to remind the user about the need to clean the filters.
Please refer to:
10-5-5-2 "Air Filter cleaning" on page 10-13
for instructions.
Table 10-7 VIVID E9 System - General Cleaning (cont’d) sheet 2 of 3
Step
Item
Description
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......