Chapter 8 - Replacement procedures
8 - 269
8-11-4-8
Removing the Graphics Adapter from the BEP5
To get access, you may remove the Hard Disk Drive (HDD).
Follow these steps to remove the HDD:
1.) Remove the screw that secures the HDD bracket. Save it for reinstallation later.
2.) Swing the HDD bracket out.
3.) Disconnect the two cables from the HDD.
Follow these steps to remove the Graphics Adapter:
1.) Disconnect the cable(s) from the Graphics Adapter.
NOTE:
Newer VIVID E9 units have a locking mechanism on the connector on the BEP’s motherboard.
2.) If a locking mechanism is installed on the connector on the BEP’s motherboard, push the plastic
slider down to release the Graphics Adapter.
3.) Unplug and remove the Graphics Adapter. Store it in an ESD safe place.
Figure 8-236 Remove fixing screw, disconnect cables
1. Fixing screw for HDD bracket.
2. Two cables to the HDD
Table 8-7
Disconnect the cables from the Graphics Adapter - BEP5
Graphics Adapter
Without DVR
With DVR
Prolink Add2 card
• Disconnect the DVI Out cable
• Disconnect the DVI Flex, Main cable
ATI FireGL V7200
• Disconnect the extra Power cable
• Disconnect the DVI Out cable
• Disconnect the extra Power cable
• Disconnect the DVI Flex, Jumper (“Z” cable)
Nvidia Quadro 2000D
and
Nvidia Quadro FX1800
• Disconnect the DVI Out cable
• Disconnect the DVI Flex, Main cable
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......