Chapter 8 - Replacement procedures
8 - 247
8-11-1-3 Back End Processor (BEP) removal procedure (cont’d)
4.) Tilt the Front Cover forward just until the Column Cover Stop Tabs clear.
5.) Lift the Lower Column Cover.
6.) Push the printer bracket toward the FEP approximately 1.3 cm (1/2 inch) to free the 3 tabs from the
BEP. Push in the direction of the white arrow in
.
7.) Remove the
BEP Cover
.
a.) Loosen the BEP Cover’s fixing screw(s).
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the
BEP Cover
away from the
BEP
.
c.) Set the
BEP Cover
aside.
8.) Remove the Printer Bracket from the BEP.
9.) Remove the two hex key screws at the inside base of the BEP.
10.)Reach inside the BEP and disconnect the
Boundary Scan Cable
from the the FEP Backplane
Connector. It is available through a cut out in the BEP’s Rear Cover.
11.)Slide the entire BEP out of chassis.
Figure 8-226 BEP hex key screws
Figure 8-227 Boundary Scan cable connects BEP to Back Plane (BEP5 is illustrated)
HEX KEY SCREWS AT
BASE
Summary of Contents for Vivid E9
Page 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Page 2: ......
Page 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Page 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Page 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Page 801: ......