GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 69
8-5-16-2 Covers under XY / Frogleg motors removal procedure (cont’d)
6.) Remove the screw securing the ground.
7.) Thread the cover(s) off the cable.
8-5-16-3
Covers under XY / Frogleg motors installation procedure
Follow these steps to install the Covers under XY / Frogleg motors:
1.) Thread the cover(s) on the cable.
2.) Perform a dry fit of the covers (confirm the covers face the correct way) before connecting the cable
and ground.
3.) Install the screw securing the ground.
4.) Connect the cable.
5.) Slide cover(s) into place.
6.) Replace the screws to the covers.
Figure 8-71 Pull down and slide XY cover out
DISCONNECT
CABLE
GROUND
SCREW
LOCATION
COVER