Section 8-10 - Front End Processor (FEP) / Card Cage parts replacement
8-10-6-3
GTX Board removal procedure
Follow these steps to remove the GTX Board:
1.) Pull out the GTX Board.
2.) Place it on an ESD safe place.
8-10-6-4
GTX Board installation procedure
Follow these steps to install the GTX Board:
NOTE:
If a GTX-TLP192 is used, it must be positioned in the right-most TX slot.
1.) Carefully align the GTX Board with the rails and push it in till it is seated in the Back Plane
connectors.
2.) Carefully install the two Front Plane Boards.
3.) Install the FEP Cover and fasten it with its fixing screws.