Chapter 4: Basic Analyzer Operation
open, wait 30 seconds for sample to drain from the
iOS
System, and then insert the vial containing the standard
or sample.
The Super
iOS
System is designed to take advantage of Sievers Standards in cartridges for automated system
protocols (see below). If you have a Super
iOS
System and want to run in grab mode, the Analyzer draws sample
from vials placed in Vial Port 1.
Figure 19: The Standard iOS System
Always stop analysis before opening the standard
iOS
System door to avoid generating erroneous
measurement data. If the flow sensor is
On
, opening the
iOS
System door while analysis is taking place causes
the flow sensor to activate and halt analysis. If the flow sensor is
Off
, opening the
iOS
System door while analysis
is taking place results in air being drawn into the Analyzer.
Warning
The
iOS
System contains two sharp needles designed to pierce the
septa of sample vials. Do not put fingers or inappropriate materials
into the
iOS
System.
Warning
Water in the standard
iOS
System may be hot. Before inserting a vial
into the standard
iOS
System, slide the door open and wait 30
seconds to allow sample to completely drain. Inserting a vial before
draining can result in potentially hot water spray projecting upward
out of the standard
iOS
System.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...