GE Analytical Instruments ©2010
61 of 226
DLM 74001-04 Rev. A
Chapter 3: Installation
Establishing a New Administrator Account for DataGuard
After you log in to the Analyzer for the first time after activating DataGuard, create a new administrator-level
account and inactivate the default administrator account to ensure Analyzer security. (Refer to Chapter 5:
Password Protection and DataGuard for complete DataGuard information.)
1. Press the
Menu
button and select the
DataGuard
tab.
2. Press the
Add User
button.
3. Specify a new User ID for the administrator and press the
Enter
button.
4. Specify the Password for the administrator User ID and press
Enter
.
5. Verify the Password and press
Enter
.
6. Press the
User Level
button and select
Administrator
.
7. Press the
Password Expired
button and select
False
.
8. Press the
Back
button.
9. Press the
Logout
button. You will now use the new administrator account and delete the default
account.
10. Press the
Login
button and enter the new User ID and Password you just created.
11. Press the
Menu
button and select the
DataGuard
tab.
12. Press the
Edit User
button.
13. Use the arrow buttons to highlight the default administrator account and press
OK
.
14. Press the
User Status
button and select
Inactive
.
15. Press the
Back
button to return to the
Menu
screen.
Setting the Clock and Time Zone
Set the Analyzer clock to ensure that all TOC measurements reflect the correct date and time:
1. Select the
Maintenance
tab.
2. Press the
Clock
button.
3. Press each date component button, enter the appropriate value, and press the
Enter
button.
4. Press the
Menu
button, press the
Advanced
button, and then press the
Advanced Setup
button.
5. (Optional) Press the
Time Zone
button, and specify the time zone via the two buttons on this screen.
Note that the time zone information you enter here does not modify the time to which you set the clock
in Step 3 above, but is merely intended to identify the time zone in exported or printed data.
•
Time Zone
— Allows you to set a text descriptor for the time zone. Usually this is a three-letter code,
such as “EST” for Eastern Standard Time or “GMT” for Greenwich Mean Time.
Summary of Contents for Sievers 500 RL
Page 8: ...GE Analytical Instruments 2010 8 of 226 DLM 74001 04 Rev A ...
Page 10: ...GE Analytical Instruments 2010 10 of 226 DLM 74001 04 Rev A ...
Page 36: ...GE Analytical Instruments 2010 36 of 220 DLM 74001 04 Rev A ...
Page 66: ...GE Analytical Instruments 2010 66 of 226 DLM 74001 04 Rev A Chapter 3 Installation ...
Page 152: ...GE Analytical Instruments 2010 152 of 226 DLM 74001 04 Rev A Chapter 7 Maintenance ...
Page 170: ...GE Analytical Instruments 2010 170 of 226 DLM 74001 04 Rev A Chapter 8 Troubleshooting ...
Page 178: ...Appendix A GE Analytical Instruments 2010 178 of 186 DLM 74001 04 Rev A ...
Page 185: ...Notes GE Analytical Instruments 2010 185 of 186 DLM 74001 04 Rev A 186 ...
Page 186: ...Notes GE Analytical Instruments 2010 186 of 186 DLM 74001 04 Rev A 186 ...